4/20/2017
Intermittent Ventricular Noncapture
What is the interpretation?
Image 1 - ECG case
Interpretation: Atrial Fibrillation, ventricular demand pacemaker (likely VVI), intermittent noncapture, adequate sensing.
Image 2 - ECG case marked
The intrinsic rhythm is atrial fibrrillation with ventricular paced beats seen pacing at about 100 bpm with adequate sensing of prior QRS. There are venrtricular paced beats that are not followed by a QRS (3,5,8,10,12,16,17 and 19). Thus, there is intermittent noncapture of the ventricles.
#676
3/07/2016
AF with failure to capture and intermittent undersening
A 75 yo with h/o DM, HTN, s/p PPM admitted for
stroke. What is wrong in the ECG?